Instruction Manual
Appendix BX Rev. 1.2
January 1997
B-2 Appendices Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
JM8
JM5
JM4
JM2
JM
1
JM7
J7
J2
SCREW
(2 PER COVER)
TERMINAL
COVERS
(PROVIDED)
EXTERNAL
EARTHING
HARDWARE
INTERNAL
EARTHING
HARDWARE
TERMINAL
STRIP
(FROM PROBE)
P00002B
TERMINAL
STRIP (FROM
ELECTRONICS)
TRANSFORMER
FRONT VIEW SIDE VIEW
Figure B-2. Heater Power Supply, Interior
Table B-1. Specifications for Heater Power Supply
Power Supply .............................................................................. 110/115/220 VAC ± 10% at 50/60 Hz
Power Requirement .................................................................... 200 VA
Humidity Range........................................................................... 95% Relative Humidity
Ambient Temperature Range...................................................... 0° to 60°C (32° to 140°F)
Vibration ...................................................................................... 5 m/sec
2
, 10 to 500 xyz plane
External Electrical Noise ............................................................. Minimum Interference
Installation Category (Overvoltage Category)............................. IEC 664 Category II
Cabling Distance Between HPS 3000 and Probe....................... Maximum 45 m (150 ft)
Cabling Distance Between HPS 3000 and CRE 3000................ Maximum 364 m (1200 ft)
Cabling Distance Between HPS 3000 and IFT 3000 .................. Maximum 364 m (1200 ft)
Approximate Shipping Weight..................................................... 20 kg (44 lbs)