Instruction Manual
106-300NFX Rev. 4.2
January 2002
2-16 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
2-3 HEATER POWER SUPPLY INSTALLATION
The Rosemount encode sheets (Prod-
uct Ordering Matrix) allow a customer
to order either the hazardous area ver-
sion of the HPS 3000 or the non-
hazardous area version. The hazard-
ous area version has the symbol
"EExd" on the apparatus nameplate.
The non-hazardous area version does
not. Ensure that if you have received
the non-hazardous version that you do
not install it in a potentially explosive
atmosphere. This also applies to the
hazardous/non-hazardous version of
the IFT 3000.
Before HPS installation, consult Safety
Data Sheet 1M03243.
a. Mechanical Installation
The outline drawing of the CENELEC ap-
proved heater power supply enclosure in
Figure 2-12, shows mounting centers and
clearances. The CENELEC approved en-
closure is designed to be mounted on a wall
or bulkhead. The heater power supply
should be installed no further than 45 m
(150 ft) from the probe. The heater power
supply must be located in a location free
from significant ambient temperature
changes and electrical noise. Ambient tem-
perature must be between 0° to 60°C (3
to 140°F).
b. Electrical Connections
1. Electrical connections should be made
as described in the electrical installa-
tion diagram, Figure 2-13. The wiring
terminals are divided into two layers:
the bottom (FROM PROBE) terminals
should be connected first, the top
(FROM ELECTRONICS) terminals
should be connected last (Figure 2-14).
Each terminal strip has a protective
cover which must be removed when
making connections. To remove the
253
(9.96)
233
(9.17)
120
(4.72)
157
(6.18)
216.0
(8.50)
264.0
(10.39)
EExd IIC T6
ENCLOSURE
14.22 (0.56) DIA
MTG HOLE (2 PLS)
NOTE: DIMENSIONS ARE IN MILLIMETERS
WITH INCHES IN PARENTHESES
UNLESS OTHERWISE INDICATED.
219005
Figure 2-12. Outline of CENELEC Approved Heater
Power Supply
terminal covers, remove two slotted
screws holding cover in place. Always
reinstall terminal covers after making
connections.
2. Power Input: 120, 220 or 240 Vac. For
120 Vac usage, install jumpers JM4
and JM1 and remove JM5 if installed.
For 220 or 240 Vac usage, install
jumper JM5 and remove JM1 and JM4
if installed (see label, Figure 2-15).
If you reconfigure the equipment for a
line voltage other than the one marked
on the serial label and the mains filter
of the power supply then you should
change the marking on the serial label
and the mains filter to state the new
line voltage.
NOTE
Fuse specifications are shown in
Figure 2-14.