4-2 LDP Operation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
key. To change the parameter, the increase and
decrease arrows are depressed until the proper
value is displayed. Depress the ENTER key to
accept the new value, or the SEL key to abort
the change. The four displays are as follows:
a. O
2
The O
2
selection will display the O
2
value on
the LED display when ENTER key is de-
pressed.
b. H
The high test gas display allows the value of
the high test gas parameter to be changed
for calibration purposes.
c. L
The low test gas display allows the value of
the low test gas parameter to be changed
for calibration purposes.
d. CAL
The calibration selection allows the operator
to initiate the calibration process when EN-
TER key is depressed.
4-4 LDP DEFAULTS
The LDP equipped IFT is programmed at the
factory with the defaults indicated in Table 4-1.
For a description of the defaults refer to Table
3-4. SETUP Sub-Menu.
4-5 CALIBRATION
a. Overview
The LDP equipped IFT is configured at the
factory for manual (semiautomatic) calibra-
tion. Calibration must be manually initiated.
Information on test gases and hardware re-
quirements may be found in paragraph
3-10, Calibration.
b. Manual Calibration
The following procedures relate to an op-
erator initiated calibration. The calibration is
manually performed by the operator upon
data queues from the IFT. Any system with-
out an MPS 3000 multiprobe test gas se-
quencer must follow these steps.
Consult Safety Data Sheet 1M03296
before performing any work on the
CENELEC approved IFT 3000.
1. Press the SEL key until CAL is shown
on the LED display.
2. Press ENTER. The CAL light is now
on. Turn on high test gas.
3. Press ENTER. The TGH light is now
flashing.
4. When the value shown on the LED
display has stabilized, press ENTER.
Turn off high test gas.
5. Turn on low test gas. Press ENTER.
The TGL light is now flashing.
6. When the value shown on the LED
display has stabilized, press ENTER.
7. Turn off low test gas. Press ENTER.
8. All indicator lights are off. Calibration
complete.