Instruction Manual
Appendix AX Rev. 2.1
February 1998
Rosemount Analytical Inc. A Division of Emerson Process Management Appendices A-9
World Class 3000
P00009A
CHECK HEATER FUSE AND TRIAC
OUTPUT TO HEATER.
HEATER MAY BE DEAD. THERMOCOUPLE
MAY BE READIN STACK TEMPERATURE.
?
HEATER, THERMOCOUPLE AND
CONTROL ARE SATISFACTORY.
REFER TO HEATER POWER
SUPPLY TROUBLESHOOTING.
IS IT HIGHER,
LOWER, OR
ZERO?
?
LOWERHIGHER
ZERO
YES
IS
HEATER
THERMOCOUPLE
CIRCUIT
OPEN?
YES
REPLACE CONTACT
AND THERMOCOUPLE
ASSEMBLY
ALLOW ADDITIONAL TIME
FOR TEMP CONTROL TO
SETTLE.
COLD JUNCTION TEMP
AT ELECTRONICS BOX
MAY BE MUCH LOWER
THAN 25°C (77°F).
2.
1.
NO
NO
CHECK THERMOCOUPLE MV
AT TERMINALS 3 AND 4 IN
PROBE JUNCTION BOX.
IS IT ABOUT SETPOINT MV ±0,2 MV?
[THE MV MEASURED IS NOT YET COLD.
JUNCTION COMPENSATED, SO IT WILL
VARY BY -0,022 MV PER DEGREE
FAHRENHEIT RISE IN THE
TEMPERATURE OF THE ELECTRONICS
BOX LOCATION. NORMAL IS SETPOINT
MV AT 25°C (77 ).]°F
3.
CONFIRM MV VALUE ON
ELECTRONICS UNIT. IF OK,
RESOLVE DIFFERENCE
BETWEEN VOLTMETER AND
ELECTRONICS.
1.
CHECK HEATER CONDITION.
TURN OFF POWER. RESISTANCE
BETWEEN TERMINALS 7 AND 8
SHOULD BE 11-14 OHMS.
RESISTANCE BETWEEN TERMINALS
6 AND 7 AND 6 AND 8 SHOULD
BE OPEN CIRCUIT. IF FAILED,
REPLACE STRUT.
2.
Figure A-9. Flowchart of Probe Related Problems, #2
AX