Instruction Manual
Appendix DX Rev. 2.1
January 1997
D-10 Appendices Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
b. Condensation Drain
To drain excess moisture from the internal
gas circuit of the MPS, periodically loosen
drain valve (46) on the bottom of pressure
regulator (50). The moisture will flow
through vinyl tubing drain (47) on the bot-
tom of pressure regulator (50) and exit the
bottom of MPS enclosure (1).
D-10 FLOWMETER ADJUSTMENTS
There are two flowmeters per flowmeter assem-
bly. The top flowmeter is factory set to 5 scfh.
The bottom flowmeter is set to 2 scfh. Should
the flow need to be changed or adjusted, use
knob on the bottom of the respective flowmeter.
D-11 ADDING PROBES TO THE MPS
This procedure is used to add a probe to the
MPS.
Disconnect and lock out power before
working on any electrical components.
a. Turn off power to the system.
b. Loosen the two captive screws holding the
MPS cover (15). Lift the cover.
c. Loosen the two captive screws that hold the
inner cover (16) and lower the cover.
d. From the backside of the inner cover, locate
the rotometer positions next to the existing
unit(s). Insert a hacksaw blade into the slots
surrounding the positions for two rotame-
ters, and saw out the knockout tabs.
e. From the front of the inner cover, install a
rotometer (P/N 771B635H01) into the top
hole and a rotameter (P/N 771B635H02)
into the bottom hole. From the backside se-
cure with brackets provided.
f. Remove four brass screw plugs (TEST GAS
IN, TEST GAS OUT, REF GAS IN, and REF
GAS OUT) for the next probe position in the
manifold.
g. Install 1/8" hose adapters (P/N
1A97553H01) into the empty holes using a
suitable pipe dope. Attach the tubing.
h. Remove a brass screw plug (P/N
1A97900H01) and install a solenoid (P/N
3D39435G01). Make sure the O-ring seals
properly.
i. Attach the hoses to the rotameter using the
existing installation as a guide. Support the
rotameter while attaching the hose.
j. Install the solenoid wire connector into the
proper position (J14-J16) on the termination
board (34).
k. Close and secure the inner cover (16) with
two captive screws. Close and secure the
outer cover (15) with two captive screws.