Instruction Manual
Appendix JX Rev. 1.0
January 1997
Rosemount Analytical Inc. A Division of Emerson Process Management Appendices J-7
World Class 3000
OPERATION
See Safety Data Sheet 1M03296 for
safety related information.
J-6 OFF-LINE AND ON-LINE OPERATIONS
The HART Communicator can be operated both
off-line and on-line. Off-line operations are those
in which the communicator is not connected to
the IFT system. Off-line operations can include
interfacing the HART Communicator with a PC
(refer to applicable HART Documentation re-
garding HART/PC applications).
In the on-line mode, the communicator is con-
nected to the 4-20 mA analog output signal line.
The communicator is connected in parallel to
the IFT, or in parallel to the 250 ohm load re-
sistor.
The opening menu (displayed on the LCD) is
different for on-line and off-line operations.
When powering up a disconnected (off-line)
communicator, the LCD will display the Main
Menu. When powering up a connected (on-line)
communicator, the LCD will display the On-line
Menu. Refer to the HART Communicator man-
ual for detailed menu information.
J-7 MENU TREE FOR HART COMMUNICATOR/
WORLD CLASS 3000 IFT APPLICATIONS
This section consists of a menu tree for the
HART Communicator. This menu is specific to
IFT 3000 applications.