2-14 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
OUTPUT JUMPER
HPS
Probe (No HPS)
Remove JM6
Install JM6
(See Figure 2-8 for jumper locations.)
Figure 2-9. IFT Microprocessor Board Jumpers
OUTPUT JUMPER
HPS
Probe (No HPS)
Remove JM1
Install JM1
Figure 2-10. Interconnect Board Jumper
Configuration
6. Before supplying power to the IFT, ver-
ify that the jumpers have been properly
set in the IFT, Figure 2-5, Figure 2-8,
and Figure 2-10.
7. Terminal strip J5 on the power supply
board is used for supplying the IFT with
power. Terminal strip J6 on the power
supply board is used to supply the
probe heater with power if an HPS is
not used (Figure 2-6).
8. Ensure that the IFT 3000 is properly
earthed by way of both the internal and
external earthing hardware.
9. Ensure that the installation does not
obscure the message on either the IFT
nameplate or the IFT lid.
c. Analog Output and Relay Output
Connections
1. The microprocessor board has a
switch to select voltage or current op-
erations. Figure 2-8 shows the switch
location. In voltage mode, output is 0-
10 V. In the current mode, the output
can be configured from the setup menu
to be 0-20 mA or 4-20 mA.
2. The analog output and relay outputs
are programmed by the user as
needed. The analog output is typically
sent to recording equipment such as
chart recorders. Relay outputs are typi-
cally sent to annunciators.
3. Relays K1 and K2 are user configur-
able from the PROBE SETUP sub-
menu (Table 3-5). Typically these are
used to indicate O
2
values above or
below specified tolerances. OK relay is
energized when unit is functioning
properly.
4. All wiring must conform to local and
national codes.
5. Connect the analog output and relay
outputs as shown in Figure 2-11.