Instruction Manual
Appendix JX Rev. 1.0
January 1997
Rosemount Analytical Inc. A Division of Emerson Process Management Appendices J-13
World Class 3000
IS
HART UNIT
INSTALLED IN
40-20 mA
SIGNAL
LOOP
?
IS
ANALOG
OUTPUT
SELECTED TO
4-20 mA
CURRENT
SIGNAL
?
IS
LOAD
RESISTANCE
250 OHMS
?
≥
IS
250 OHM
LOAD
RESISTOR
CORRECTLY
INSTALLED (SEE
INSTALLATION)
?
IS
HART UNIT
CONFIGURED
FOR IFT
?
IS
HART UNIT
INSTALLED
PARALLEL
TO IFT
?
RECONFIGURE HART
UNIT FOR IFT.
YES
YES
YES
YES
YES
YES
PROGRAM IFT FOR
4-20 mA OUTPUT SIGNAL.
SELECT CURRENT MODE
USING CURRENT/VOLTAGE
SELECTION SWITCH ON
IFT MICROPROCESSOR
BOARD.
CONNECT THE HART UNIT
TO THE 4-20 mA SIGNAL
LOOP. REFER TO INSTALLATION.
IF LOAD RESISTANCE IS
<250 OHMS. THE 250 OHM
LOAD RESISTOR MUST BE
USED.
INSTALL 250 OHM LOAD
RESISTOR. (SEE
INSTALLATION).
CONTACT SERVICE
REPRESENTATIVE.
USE LEAD SET TO
CONNECT HART UNIT IN
PARALLEL TO IFT. HART
UNIT MUST BE
CONNECTED TO 4-20 mA
SIGNAL LINE. (SEE
INSTALLATION).
VERIFY PROPER IFT
CONFIGURATION.
CONSULT MENU TREE
(FIGURE J-4 OF THIS
APPENDIX). REFER TO
DEVICE CONFIGURATION
SUB-MENU.
NO
NO
NO
NO
NO
(CONTINUED FROM
SHEET 1 OF 2)
NO
35870007
Figure J-5. Model Number 275D9E, Troubleshooting Flowchart (Sheet 2 of 2)