E-8 Appendices Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
Table E-2. GUI Equipped IFT Fault Finding
SYMPTOM COMPONENT FAILURE
1. Display is blank. Possible failure within IFT. Check LED on microprocessor board.
2. C Err is displayed. Repeat calibration sequence. If error persists, troubleshoot major
components.
3. H Err is displayed. Ensure jumpers are set correctly on IFT. If system is equipped with HPS
refer to Appendix BX, Troubleshooting, for additional troubleshooting
procedures.
4. TGLow is displayed. Possible failure within the MPS. Refer to Appendix DX, Troubleshooting,
for additional troubleshooting procedures.
*5. HiO
2
is displayed. Possible failure within the probe. Ensure the high alarm level has been en-
tered correctly. Refer to Appendix AX, Troubleshooting, for additional
troubleshooting procedures.
*6. LoO
2
is displayed. Possible failure within the probe. Ensure the low alarm parameter has been
entered correctly. Refer to Appendix AX, Troubleshooting, for additional
troubleshooting procedures.
7. R Hi is displayed. Cell resistance has exceeded upper limit. Ensure resistance value has been
entered correctly. Refer to Appendix AX, Troubleshooting, for additional
troubleshooting procedures.
8. Off The probe has been turned OFF because the IFT cannot control the heater
temperature.
9. PRBE The probe is disconnected, cold, or leads are reversed.
*HiO
2
and LoO
2
can occur in the system without system failure.