Vista de pagina 34
Instruction Manual
106-300NFX Rev. 4.2
January 2002
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-17
World Class 3000
+
+
+
+
+
+
+
+
+
-
-
-
-
-
-
-
-
-
BK
WH
J9
J8
J3
J2 J1
TRIAC RELAY
STACK
TC
ANALOG
HEATER
PROBE
MV
PROBE
MV
PROBE
TC
PROBE
TC
AD590
SM. BR CELL+
RD HTR TC +
WH CELL -
GN HTR TC -
LG. BR
BL
GN/YE
STACK
TC
PROBE
HEATER
RH
N
L
LINE
VOLTAGE
SHIELD
TOP
BOTTOM
4 TWISTED PAIR SHIELDED
#22 AWG BY CUSTOMER
2 TWISTED PAIR SHIELDED
#22 AWG BY CUSTOMER
(OPTIONAL)
HPS 3000 INTERFACE MODULE
LINE
WC PROBE 3000 CENELEC APPROVED
12345678
WH
SM. BR
RD
GN
PROBE INTERIOR
GN/YE
BL
LG. BR
GN CELL -VE
OR CELL +VE
YE CHROMEL
RD ALUMEL
GN
BK
BK
}
HEATER
PROBE JUNCTION
BOX WIRING
ALWAYS DISCONNECT LINE VOLTAGE
FROM HEATER POWER SUPPLY AND
ANALOG ELECTRONICS (IF USED)
BEFORE CHANGING JUMPERS.
JUMPER
CONFIGURATIONS
LINE VOLTAGE
SELECTION
JUMPER
(INSTALL)
JUMPER
(INSTALL)
HEATER
POWER
JUMPER
120 V.A.C.
220/240 V.A.C.
JM4, JM1
JM5
REMOTE
ON
REMOVE JM2
INSTALL JM2
ELECTRONICS
SELECTION
JUMPER
PROBE HEATER
VOLTAGE SECTION
REMOVE JM3, JM6
NEW GENERATION
ELECTRONICS
WORLD CLASS PROBE JM7
RELAY WIRE IS OPTIONAL; RELAY CAN BE PERMANENTLY
ENABLED WITH JUMPER IF NOT USED
STACK TC WIRING AS REQUIRED
ALL WIRES #16-#22 AWG TWISTED PAIR WITH SHIELD
BY CUSTOMER EXCEPT AS NOTED
SPECIAL PROBE CABLE BETWEEN PROBE AND HPS
BY ROSEMOUNT
REMOVE JM1 ON INTERCONNECT BOARD (IFT 3000)
REMOVE JM6 ON MICROPROCESSOR BOARD
IF RELAY WIRE OF NOTE 1 INSTALLED THEN REMOVE
JM2 ON HPS 3000
IF STACK TEMPERATURE NOT USED
IF MPS 3000 NOT USED
1 RELAY PER PROBE AVAILABLE FOR CALIBRATION
STATUS INDICATION (48 V max, 100 mA max)
CABLE COLORS SHOWN HERE APPLY TO ROSEMOUNT
SUPPLIED SPECIAL CABLE FITTED WITH EExd GLANDS
(P/N 1U03066)
NOTES
35870008
B
A
Figure 2-13. Wiring layout for IFT 3000 (CENELEC approved) with HPS (Sheet 1 of 2)