2-8 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
c. Reference Air Package
After the oxygen analyzing (probe) unit is
installed, connect the reference gas air set
to the probe cover. The reference gas air
set should be installed in accordance with
Figure 2-3.
d. Service Required.
1. Power input: 44 VAC from HPS 3000
or IFT 3000.
2. Compressed air: 68.95 kPa (10 psig)
minimum, 1551.38 kPa (225 psig)
maximum at 56.6 L/hr (2 scfh) maxi-
mum; supplied by one of the following
(less than 40 parts-per-million total hy-
drocarbons).
(a) Instrument air - clean, dry.
(b) Bottled standard air with step-down
regulator.
(c) Bottled compressed gas mixture
(20.95% oxygen in nitrogen).
(d) Other equivalent clean, dry, oil-free
air supply.
17300016
TO PROBE
JUNCTION BOX
REF GAS SET
263C152G01
1 FLOWMETER 0.2-2.0 SCFH 771B635H02
2 2" PRESSURE GAGE 0-15 PSIG 275431-006
3 COMBINATION FILTER-REG. 0-30 PSIG 4505C21G01
NOTE: DIMENSIONS ARE IN MILLIMETERS
WITH INCHES IN PARENTHESES.
1
2
3
122.17 (4.81)
FLOW SET
POINT KNOB
0.125-27 NPT FEMALE
OUTLET CONNECTION
30.22
(1.19)
254 REF
(10.0)
DRAIN VALVE
79.25 (3.12) MAX
215.90 MAX
(8.50)
50.80
(2.0)
2 MOUNTING HOLES
81.03 (3.19) LG
THROUGH BODY FOR
7.92 (0.312) DIA BOLTS
38.10
(1.50)
57.15 (2.250)
SCHEMATIC HOOKUP FOR REFERENCE AIR SUPPLY ON OXYGEN ANALYZER PROBE HEAD.
OUTLET
0.25-18 NPT FEMALE
INLET CONNECTION
COMPRESSED AIR SUPPLY
10-225 PSIG MAX PRESSURE
6 (0.250) OD TUBING
(SUPPLIED BY CUSTOMER)
6 (0.250) OD
TUBE COMPRESSION
FITTING (SUPPLIED BY WECO)
NOTE: DIMENSIONS ARE IN MILLIMETERS
WITH INCHES IN PARENTHESES.
Figure 2-3. Air set, Plant Air Connection