Instruction Manual
106-300NFX Rev. 4.2
January 2002
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-7
World Class 3000
4. If using the optional ceramic diffuser
element, the vee deflector must be cor-
rectly oriented. Before inserting the
probe, check the direction of flow of the
gas in the duct. Orient the vee deflector
on the probe so that the apex points
upstream toward the flow (Figure 2-2).
This may be done by loosening the
setscrews, and rotating the vee de-
flector to the desired position.
Retighten the setscrews.
5. In horizontal installations, the probe
cover should be oriented so that the
system cable drops vertically from the
probe cover. In a vertical installation,
the system cable can be oriented in
any direction.
6. If the system has an abrasive shield,
check the diffusion element dust seal
packings. The joints in the two pack-
ings must be staggered 180°. Also,
make sure that the packings are in the
hub grooves as the probe slides into
the 15° forcing cone in the abrasive
shield.
NOTE
If process temperatures will exceed
1000°F (538°C), use anti-seize com-
pound on stud threads to ease future
removal of probe.
APEX
FILTER
GAS FLOW
DIRECTION
VEE
DEFLECTOR
CERAMIC
DIFFUSION
ELEMENT
VEE
DEFLECTOR
SETSCREW
23800008
Figure 2-2. Orienting the Optional Vee Deflector
7. Insert the probe through the opening in
the mounting flange and bolt the unit to
the flange.
8. Ensure that probe is properly earthed
by way of both the internal and external
points.
9. Ensure that the installation does not
obscure the messages on either the
probe nameplate or the junction box
lid.