Instruction Manual
Appendix DX Rev. 2.1
January 1997
D-2 Appendices Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
TUBE
35870005
ROTOMETER
POWER
SUPPLY
REGULATOR
SOLENOID
MANIFOLD
HOSE
ADAPTER
TERMINATION
BOARD
CABLE
GRIP
Figure D-2. Multiprobe Test Gas Sequencer, Interior
Table D-1. Specifications for Multiprobe Test Gas Sequencer
Power Supply ........................................................................................... 110/115/220 Vac ±10% at 50/60 Hz
Power Requirement ................................................................................. 15 VA (Watts)
Electrical Classification ............................................................................ NEMA 4X (IP56)
Humidity Range........................................................................................ 95% Relative Humidity
Ambient Temperature Range................................................................... -30° to 71°C (-20° to 160°F)
Vibration ................................................................................................... 5 m/sec
2
, 10 to 500 xyz plane
External Electrical Noise .......................................................................... Minimum Interference
Installation Category (Overvoltage Category).......................................... IEC 664 Category II
Piping Distance Between MPS 3000 and Probe ..................................... Maximum 91 m (300 ft)
Cabling Distance Between MPS 3000 and Electronics Package ............ Maximum 303 m (1000 ft)
In Calibration Status Relay....................................................................... 48V max, 100 mA max
Cabling Distance Between MPS 3000 and Status Relay Indicator ......... Maximum 303 m (1000 ft)
Approximate Shipping Weight.................................................................. 16 kg (35 lbs)