Instruction Manual
106-300NFX Rev. 4.2
January 2002
2-6 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
PROBE MOUNTING
ABRASIVE SHIELD MOUNTING
INSTALLATION FOR MASONRY
WALL STACK CONSTRUCTION
INSTALLATION FOR METAL
WALL STACK OR DUCT
CONSTRUCTION
13 (0.50)
95 (3.75)
MIN DIA HOLE
IN WALL
STACK OR DUCT
METAL WALL
MTG HOLES
SHOWN ROTATED
22.5 OUT OF
TRUE POSITION
o
WELD OR BOLT
MOUNTING PLATE TO
METAL WALL OF STACK
OR DUCT. JOINT MUST
BE AIR TIGHT.
13 (0.50)
114 (4.50)
O.D. REF
PIPE 4.00 SCHED 40
PIPE SLEEVE (NOT
BY ROSEMOUNT)
LENGTH BY CUSTOMER
MASONRY
STACK WALL
OUTSIDE WALL
SURFACE
JOINT MUST
BE AIRTIGHT
MTG HOLES
SHOWN ROTATED
22.5 OUT OF
TRUE POSITION
o
FIELD WELD
PIPE TO
ADAPTOR PLATE
BOLT ADAPTOR
PLATE TO OUTSIDE
WALL SURFACE
NOTE: ALL MASONRY STACK WORK AND JOINTS EXCEPT
MOUNTING PLATE NOT FURNISHED BY ROSEMOUNT.
82.5 (3.25)
MIN DIA HOLE
IN WALL
STACK OR DUCT
METAL WALL
WELD OR BOLT MOUNTING
PLATE TO METAL WALL
OF STACK OR DUCT.
JOINT MUST BE AIR TIGHT.
FIELD WELD
PIPE TO
ADAPTOR PLATE
102 (4.0)
O.D. REF
PIPE 3.5 SCHED 40
PIPE SLEEVE (NOT
BY ROSEMOUNT)
LENGTH BY CUSTOMER
MASONRY
STACK WALL
OUTSIDE WALL
SURFACE
JOINT MUST
BE AIRTIGHT
BOLT MOUNTING
PLATE TO OUTSIDE
WALL SURFACE
P00007
Figure 2-1. Probe Installation (Sheet 5 of 5)