Instruction Manual
106-300NFX Rev. 4.2
January 2002
Rosemount Analytical Inc. A Division of Emerson Process Management Setup and Operation 3-3
World Class 3000
3-4 STATUS LINE
The top line of the LCD display (3, Figure 3-1) is
a status line that always displays system status,
menu name, and O
2
level. The system status
displays are:
a. OK - System is functioning correctly.
b. CAL - Calibration in progress.
c. C Err - Calibration error.
d. H Err - Heater error.
e. TGLow - Test gas is low.
f. HiO
2
- O
2
value is above the high alarm
limit.
g. LoO
2
- O
2
value is below the low alarm limit.
h. R Hi - Resistance is above the high limit.
i. Off - The probe has been turned off be-
cause the IFT cannot control the heater
temperature.
j. PRBE - The probe is disconnected, cold, or
leads are reversed.
3-5 QUICK REFERENCE CHART
The quick reference chart on pages 3-4 and 3-5
is designed to help you determine how to get
where you want to be in the menu system. The
chart shows all the available menu and sub-
menu options for the IFT. Follow the lines to
determine which choices to make. Moving down
a level on the chart is accomplished by use of
the ENTER key. To move up a level on the
chart, press the ESCAPE key.
3-6 MAIN MENU
When power is first applied to the IFT, the MAIN
menu (Table 3-2) is initially displayed. It is from
the MAIN menu that the PROBE DATA (Table
3-3), CALIBRATE O
2
(Table 3-4), and SETUP
(Table 3-5) menus can be accessed.
Table 3-2. Main Menu
MENU SELECTION DESCRIPTION
PROBE DATA Refer to Table 3-3.
CALIBRATE O
2
Refer to Table 3-4.
SETUP Refer to Table 3-5.
Table 3-3. PROBE DATA Sub-Menu.
SUB-MENU
SELECTION PARAMETER DESCRIPTION
Process Data O
2
__%
Efficiency ENA/DIS
Stack Temp __DegC
O
2
value for the probe.
Enable/Disable efficiency display.
Stack temperature.
Diagnostic Data
Temperature Cell __DegC
Stack __DegC
Cold Junct __DegC
Cell temperature of the probe.
Stack temperature.
Cold junction temperature.
Voltages Cell __mV
Cell T/C __mV
Stk T/C __mV
Cold Jnt __mV
Cell voltage of the probe.
Cell thermocouple voltage of the probe.
Stack thermocouple voltage.
Cold junction voltage.
Output Values Analog __% FS
K1 State OFF/ON
K2 State OFF/ON
Analog output voltage.
Status of relay 1.
Status of relay 2
3