Instruction Manual
106-300NFX Rev. 4.2
January 2002
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-19
World Class 3000
JM8
JM5
JM4
JM2
JM
1
JM7
J7
J2
SCREW
(2 PER COVER)
TERMINAL
COVERS
(PROVIDED)
EXTERNAL
EARTHING
HARDWARE
INTERNAL
EARTHING
HARDWARE
TERMINAL
STRIP
(FROM PROBE)
219006-1
TERMINAL
STRIP (FROM
ELECTRONICS)
TRANSFORMER
FRONT VIEW SIDE VIEW
FUSES
FUSE
NOTE: FUSES SHOWN (F1 THROUGH F4) ARE 5 AMP, ANTI-SURGE,
TYPE T TO IEC127 (ROSEMOUNT PART NUMBER 1L01293H02).
Figure 2-14. CENELEC Approved Heater Power Supply Wiring Connections
ALWAYS DISCONNECT LINE VOLTAGE
FROM HEATER POWER SUPPLY AND
ANALOG ELECTRONICS (IF USED)
BEFORE CHANGING JUMPERS.
JUMPER
CONFIGURATIONS
LINE VOLTAGE
SELECTION
JUMPER
(INSTALL)
JUMPER
(INSTALL)
HEATER
POWER
JUMPER
120 V.A.C.
220/240 V.A.C.
JM4, JM1
JM5
*ON
REMOTE
INSTALL JM2
REMOVE JM2
ELECTRONICS
SELECTION
JUMPER
PROBE HEATER
VOLTAGE SELECTION
REMOVE JM3, JM6
NEW GENERATION
ELECTRONICS
*WORLD CLASS PROBE
(44V)
JM7
219007
Figure 2-15. Jumper Selection Label.
If you reconfigure the equipment for a line voltage other than the one marked on the serial la-
bel and the mains filter of the power supply then you should change the marking on the serial
label and the mains filter to state the new line voltage.