January 2002
5-2 Troubleshooting Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
5-3 SYSTEM TROUBLESHOOTING
The IFT provides system failure information with
two different error message formats. The error
messages vary due to system configuration.
Refer to Appendix EX, IFT 3000 Troubleshoot-
ing.
a. GUI Equipped IFT
The status line of the GUI equipped IFT will
display one of ten conditions: OK, CAL
(calibration), C Err (calibration error), H Err
(heater error), TGLow (test gas low), HiO
2
(high O
2
level), LoO
2
(low O
2
level), and
R Hi (high resistance level), Off and PRBE.
Refer to Table E-2, Appendix EX for addi-
tional troubleshooting information on the
GUI equipped IFT.
Consult Safety Data Sheet 1M03296
before performing any work on the
CENELEC approved IFT 3000.
1. OK - The system is operating
normally.
2. CAL - The system is currently
undergoing calibration.
3. C Err - An error was detected during
the calibration process.
4. H Err - There is a fault within the heater
system.
5. TGLow - There is no test gas pressure.
6. HiO
2
- The O
2
value is above the high
alarm limit.
7. LoO
2
- The O
2
value is below the low
alarm limit.
8. R Hi - The cell resistance is above the
high limit.
9. Off - The probe has been turned off
because the IFT cannot control the
heater temperature.
10. PRBE - The probe is disconnected,
cold, or leads are reversed.