TROUBLESHOOTING
D-3 OVERVIEW
This section describes troubleshooting for the
Multiprobe Test Gas Sequencer. Additional
troubleshooting information can be found in the
Instruction Bulletin for the electronics package.
Install all protective equipment covers
and safety ground leads after trouble-
shooting. Failure to replace covers
and ground leads could result in seri-
ous injury or death.
D-4 TROUBLESHOOTING
Table D-2 provides a guide to fault finding fail-
ures within the MPS. The flowchart in Figure
D-5 provides an alternate approach to fault
finding MPS related problems.
Table D-2. Fault Finding
SYMPTOM CHECK FAULT REMEDY
1. Power to solenoid,
test gas not
released to probe.
Test gas
Solenoid
Insufficient test gas
Solenoid failure
Install new test gas
tanks.
Replace solenoid.
2. No power to
solenoid.
Power supply output Power supply failure Replace power supply.
Fuses in power supply Fuse blown Replace fuse.
Main power source Main power off Reestablish power.