Reference Manual
00809-0200-4801, Rev CA
July 2010
6-13
Rosemount 3051S
Figure 6-9. Example of PID
Control within a Pressure
Transmitter
Using SPM, it is possible to detect problems in the control loop. For example,
an increase in the cycling or oscillation of the control loop could be indicated
by an increase in the standard deviation.
If SPM has already been enabled, it must first be disabled, before any
additional SPM blocks can be configured. First, set
SPM_Active = Disabled
Next, to configure SPM2 to monitor the PID control loop, you would set the
parameters as shown in Figure 6-10.
SPM2_Block_Tag = FFPID_RMT5
SPM2_Block_Type = PID Block
SPM2_Parameter_Index = OUT
Figure 6-10. Example
Configuration of SPM Using a
Function Block
Set the High Variation Limit (or other limits) as desired, and set the user
command to “Learn” in order to begin the learning process for this SPM Block.
Finally, verify that SPM_Monitoring_Cycle and SPM_Bypass_Verification are
set as desired, and set the SPM_Active Parameter to “Enabled” (with or
without the filter)