Reference Manual
00809-0200-4801, Rev CA
July 2010
A-15
Rosemount 3051S
ADVANCED
DIAGNOSTICS
TRANSDUCER BLOCK
(ADB)
Table A-5. Advanced Diagnostic
block Parameters
Parameter Index Valid Range Initial Value Units Other Description
ST_REV 1 0-255 Read
Only
The revision level of the static data associated with the
function block.
TAG_DESC 2 spaces N/A The user description of the intended application of the block.
STRATEGY 3 0 N/A The strategy field can be used to identify grouping of blocks.
ALERT_KEY 4 1-255 0 N/A The identification number of the plant unit.
MODE_BLK 5 O/S N/A The actual, target, permitted, and normal modes of the block.
BLOCK_ERR 6 E Read
Only
This parameter reflects the error status associated with the
hardware or software components associated with a block. it is
a bit string, so that multiple errors may be shown.
UPDATE_EVT
NOT
IMPLEMENTED
7 N/A Read
Only
This alert is generated by any change to the staic data.
BLOCK_ALM
NOT
IMPLEMENTED
8 N/A The BLOCK_ALM is used for all configuration, hardware,
connection failure or system problems in the block. The cause
of the alert is entered in the subcode field. The first alert to
become active will set the Active status in the Status attribute.
As soon as the Unreported status is cleared by the alert
reporting task, another block alert may be reported without
clearing the Active status, if the subcode has changed.
TRANSDUCER_
DIRCTORY
9 N/A Read
Only
A directory that specifies the number and starting indicies of
the transducers in the transducer block.
TRANSDUCER_
TYPE
10 100 E Read
Only
Identifies the transducer that follows.
XD_ERROR 11 E Read
Only
Provides additional error codes related to transducer blocks.
COLLECTION_
DIRECTORY
12 N/A Read
Only
A directory that specifies the number, starting indicies, and DD
Item ID's of the data collections in each transducer block.
ADB_STATUS 13 Quality:
Good
SubStatus:
Non-Specifi
c
Limit: 0
N/A Read
Only
Status of Advanced Diagnostics Block.
DIAG_EVT 14 Unack: 0
Alarm
State: 0
Time
Stamp:0
Subcode: 0
Value: 0
N/A Read
Only
Diagnostic event
SPM_ACTIVE 15 0 = Disabled
Oxfe = Enabled
with HP filter
0xff = Enabled
without filter
Disabled N/A Enables/Disables the Statistical Process Monitoring algorithm.
Allows for activation of a high-pass filter for standard deviation
calculations.
SPM_
MONITORING_
CYCLE
16 1 to 1440 15 min. Time Perioed for Mean and Stdev calculations during Learning
and Monitoring