Reference Manual
00809-0200-4801, Rev CA
July 2010
Rosemount 3051S
B-10
FUNCTIONAL
SPECIFICATIONS
Range and Sensor Limits
Transmitter with Coplanar Sensor Module (Single Variable)
Transmitter with In-Line Sensor Module
Transmitter with MultiVariable Sensor Module (3051SMV_ _1, 3051SMV_ _2, 3051SF_1, 3051SF_2,
3051SF_5, and 3051SF_6)
Range
DP Sensor
(1)
(3051S_CD, 3051SMV_ _3, 4, or D
3051SF_3, 4, or 7, 3051S_LD)
(1) 3051SF flowmeters only available with ranges 1, 2, and 3.
GP Sensor
(3051S_CG, 3051S_LG)
AP Sensor
(3051S_CA, 3051S_LA)
Lower (LRL)
(2)
(2) The Lower Range Limit (LRL) is 0 inH
2
0 (0 mbar) for Ultra for Flow performance class and 3051SF flowmeters.
Upper (URL) Lower (LRL)
(3)
(3) Assumes atmospheric pressure of 14.7 psig (1 bar).
Upper (URL) Lower (LRL) Upper (URL)
0 -3 inH
2
O
(-7,5 mbar)
3 inH
2
O
(7,5 mbar)
N/A N/A 0 psia
(0 bar)
5 psia
(0,34 bar)
1 -25 inH
2
O
(-62,3 mbar)
25 inH
2
O
(62,3 mbar)
-25 inH
2
O
(-62,3 mbar)
25 inH
2
O
(62,3 mbar)
0 psia
(0 bar)
30 psia
(2,07 bar)
2 -250 inH
2
O
(-0,62 bar)
250 inH
2
O
(0,62 bar)
-250 inH
2
O
(-0,62 bar)
250 inH
2
O
(0,62 bar)
0 psia
(0 bar)
150 psia
(10,34 bar)
3 -1000 inH
2
O
(-2,49 bar)
1000 inH
2
O
(2,49 bar)
-1000 inH
2
O
(-2,49 bar)
1000 inH
2
O
(2,49 bar)
0 psia
(0 bar)
800 psia
(55,16 bar)
4 -300 psi
(-20,7 bar)
300 psi
(20,7 bar)
-300 psi
(-20,7 bar)
300 psi
(20,7 bar)
0 psia
(0 bar)
4000 psia
(275,8 bar)
5 -2000 psi
(-137,9 bar)
2000 psi
(137,9 bar)
-2000 psi
(-137,9 bar)
2000 psi
(137,9 bar)
N/A N/A
Range
GP Sensor
(3051S_TG)
AP Sensor
(3051S_TA)
Lower (LRL)
(1)
Upper (URL) Lower (LRL) Upper (URL)
1 -14.7 psig (-1,01 bar) 30 psig (2,07 bar) 0 psia (0 bar) 30 psia (2,07 bar)
2 -14.7 psig (-1,01 bar) 150 psig (10,34 bar) 0 psia (0 bar) 150 psia (10,34 bar)
3 -14.7 psig (-1,01 bar) 800 psig (55,16 bar) 0 psia (0 bar) 800 psia (55,16 bar)
4 -14.7 psig (-1,01 bar) 4000 psig (275,8 bar) 0 psia (0 bar) 4000 psia (275,8 bar)
5 -14.7 psig (-1,01 bar) 10000 psig (689,5 bar) 0 psia (0 bar) 10000 psia (689,5 bar)
(1) Assumes atmospheric pressure of 14.7 psig (1 bar).
Range
DP Sensor
Lower (LRL)
(1)
(1) Lower (LRL) is 0 inH
2
O (0 mbar) for Ultra for Flow and 3051SF_ Flowmeters.
Upper (URL)
1 -25.0 inH
2
O (-62,3 mbar) 25.0 inH
2
O (62,3 mbar)
2 -250.0 inH
2
O (-0,62 bar) 250.0 inH
2
O (0,62 bar)
3 -1000.0 inH
2
O (-2,49 bar) 1000.0 inH
2
O (2,49 bar)
Range
Static Pressure Sensor (GP/AP)
Lower (LRL) Upper (URL)
(1)
3 GP
(2)
: -14.2 psig (0,98 bar)
AP: 0.5 psia (34,5 mbar)
GP: 800 psig (55,16 bar)
AP: 800 psia (55,16 bar)
4 GP
(2)
: -14.2 psig (0,98 bar)
AP: 0.5 psia (34,5 mbar)
GP: 3626 psig (250 bar)
AP: 3626 psia (250 bar)
(1) For SP Range 4 with DP Range 1, the URL is 2000 psi (137,9 bar).
(2) Inert Fill: Minimum pressure = 1.5 psia (0,10 bar) or -13.2 psig (-0,91 bar).