Rosemount 3051S Series
Draft Range Installation
For the 3051S_CD0 draft range pressure transmitter, it is best to mount the
transmitter with the isolators parallel to the ground. Installing the transmitter in
this way reduces oil mounting effect and provides for optimal temperature
performance.
Be sure the transmitter is securely mounted. Tilting of the transmitter may
cause a zero shift in the transmitter output.
Reducing Process Noise
There are two recommended methods of reducing process noise: output
damping and, in gage applications, reference side filtering.
Reference Side Filtering
In gage applications it is important to minimize fluctuations in atmospheric
pressure to which the low side isolator is exposed. One method of reducing
fluctuations in atmospheric pressure is to attach a length of tubing to the
reference side of the transmitter to act as a pressure buffer.
Another method is to plumb the reference side to a chamber that has a small
vent to atmosphere. If multiple draft transmitters are being used in an
application, the reference side of each device can be plumbed to a chamber
to achieve a common gage reference.
Environmental Access requirements and cover installation on page 2-4 can help optimize
transmitter performance. Mount the transmitter to minimize ambient
temperature changes, vibration, mechanical shock, and to avoid external
contact with corrosive materials. Appendix B: Specifications and Reference
Data lists temperature operating limits.