
Transmitter Configuration and Use: Series 1000 and 2000 Transmitters 79
Required Configuration Optional ConfigurationUsing the TransmitterFlowmeter Startup
Chapter 8
Optional Configuration
8.1 Overview
This chapter describes transmitter configuration parameters that may or may not be used, depending
on your application requirements. For required transmitter configuration, see Chapter 6.
8.2 Configuration map
Different transmitters support different parameters and features. Additionally, different configuration
tools allow you to configure different features. Table 8-1 lists the optional configuration parameters.
For each parameter, the table also lists the transmitters that support that parameter and a reference to
the section where the parameter is discussed.
The configuration map uses the following codes:
•
AN – transmitters with the analog outputs option board
•
IS – transmitters with the intrinsically safe outputs option board
•
CIO – transmitters with the configurable input/outputs option board
Default values and ranges for the most commonly used parameters are provided in Appendix A.
8.3 How to access a parameter for configuration
In general, all parameters discussed in this chapter can be configured either with ProLink II or the
Communicator, but cannot be configured with the display. Exceptions are noted in the configuration
map.
For information on the menu structure for each transmitter, and how to access a particular parameter,
see the appendix for your transmitter, as listed below. Within that appendix, refer to the menu
flowcharts for the communication tool you are using: ProLink II, the Communicator, or the display.
• Model 1500 AN – see Appendix E
• Model 2500 CIO – see Appendix F
• Model 1700/2700 AN – see Appendix G
• Model 1700/2700 IS – see Appendix H
• Model 2700 CIO – see Appendix I
Note: If you are using AMS, the AMS interface will be similar to the ProLink II interface.
Comentarios a estos manuales