Transmitter Configuration and Use: Series 1000 and 2000 Transmitters 77
Using the Transmitter continued
Required Configuration Optional ConfigurationUsing the TransmitterFlowmeter Startup
With ProLink II software
Table 7-5 shows how you can control the totalizers and inventories using ProLink II software.
To get to the Totalizer Control screen:
1. Click
ProLink.
2. Select
Totalizer Control.
Reset mass totalizer
(1)
Scroll until the mass totalizer value appears.
Select.
If a password is required (CODE? appears), enter it and Select.
(2)
Scroll until RESET appears beneath the current totalizer value.
Select (
YES alternates with RESET).
Select (mass totalizer resets).
Reset volume totalizer
(1)
Scroll until the volume totalizer value appears.
Select.
If a password is required (
CODE? appears), enter it and Select.
(2)
Scroll until RESET appears beneath the current totalizer value.
Select (
YES alternates with RESET).
Select (volume totalizer resets).
Reset API totalizer
(1)
Scroll until an API totalizer value appears.
Select.
A password is always required (CODE? appears). Enter it and Select.
(2)
Scroll until RESET appears beneath the current totalizer value.
Select (
YES alternates with RESET).
Select (API totalizer resets).
(1) With rev3.3 or higher of the transmitter software, this feature may be enabled or disabled. See Section 8.14.1.
(2) A password is required only if the petroleum measurement application (API feature) is enabled. If the petroleum measurement
application is not enabled, no password is required even if the display password has been enabled.
Table 7-5 Totalizer and inventory control with ProLink II software
To accomplish this On the totalizer control screen...
Stop all totalizers and inventories (mass, volume, and API) Click Stop
Start all totalizers and inventories (mass, volume, and API) Click Start
Reset mass totalizer Click Reset Mass Total
Reset volume totalizer Click
Reset Volume Total
Simultaneously reset all totalizers (mass, volume, and API) Click Reset
Simultaneously reset all inventories (mass, volume, and API)
(1)
(1) If enabled in the ProLink II preferences.
Click
Reset Inventories
Table 7-4 Totalizer and inventory control with the display (continued)
To accomplish this Press this sequence of buttons