22 Transmitter Configuration and Use: Series 1000 and 2000 Transmitters
Connecting with ProLink II Software continued
Follow the instructions below to make the connection.
1. Connect the HART interface to your PC’s serial port. Then connect the leads of the HART
interface to terminals 21 and 22.
2. Add 250–600 resistance to the connection, as required.
3. Start ProLink II software. From the Connection menu, click on
Connect to Device.
4. In the screen that appears, set
Protocol to HART Bell 202. Baud rate, Stop bits, and Parity
are automatically set to the values required by HART protocol. Specify the remaining
connection parameters as shown in Table 3-3.
5. Click the
Connect button. ProLink will attempt to make the connection.
6. If an error message appears:
a. Ensure that you are using the correct COM port.
b. Check all the wiring between the PC and the transmitter.
c. Increase or decrease the resistance.
Table 3-3 HART connection parameters for ProLink II
Connection parameter HART setting
Address/Tag Configured HART polling address (default = 0)
COM port COM port assigned to PC serial port