Transmitter Configuration and Use: Series 1000 and 2000 Transmitters 39
Flowmeter Startup continued
Required Configuration Optional ConfigurationUsing the TransmitterFlowmeter Startup
5.5.2 Zeroing with the display
See Figure 5-7 for the flowmeter zero procedure.
Note the following:
If the off-line menu has been disabled, you will not be able to zero the
transmitter with the display. For information about enabling and disabling
the off-line menu, see Section 8.14.1.
You cannot change the zero time with the display. If you need to change the
zero time, you must use the Communicator or ProLink II software.
Figure 5-7 Display – Flowmeter zero procedure
Model 1700 AN
Model 1700 IS
Model 2700 AN
Model 2700 IS
Model 2700 CIO
TEST FAIL
ZERO and YES?
alternate on display
TEST OK
Troubleshoot
Select
OFF-LINE MAINT
Scroll and Select
simultaneously for 4 seconds
OFF-LINE ZERO
Scroll
Scroll
Select
Select
Dots traverse top line
of display
Exit
Select