114 Transmitter Configuration and Use: Series 1000 and 2000 Transmitters
Troubleshooting continued
mA output < 4 mA Process condition below LRV Verify process.
Change the LRV. See Section 6.5.2.
Fault condition if fault indicator is set to
internal zero
Check the fault indicator settings to verify
whether or not the transmitter is in a fault
condition. See Section 6.5.4.
If a fault condition is present, see Section 10.7.
Open in wiring Verify all connections.
Bad mA receiving device Check the mA receiving device or try another
mA receiving device. See Section 10.15.
Channel not configured for mA
operation (CIO transmitters only)
Verify channel configuration.
Bad output circuit Measure DC voltage across output to verify that
output is active.
Output not powered (IS transmitters
only)
Check transmitter wiring. See Appendix B for
diagrams.
No frequency output Process condition below cutoff Verify process.
Change the cutoff. See Section 8.6.
Fault condition if fault indicator is set to
downscale or internal zero
Check the fault indicator settings to verify
whether or not the transmitter is in a fault
condition. See Section 6.6.6.
If a fault condition is present, see Section 10.7.
Slug flow See Section 10.16.
Flow in reverse direction from
configured flow direction parameter
Verify process.
Check flow direction parameter. See
Section 8.9.
Verify sensor orientation. Ensure that flow
direction arrow on sensor case matches process
flow.
Bad frequency receiving device Check the frequency receiving device or try
another frequency receiving device. See
Section 10.15.
Incorrect terminal configuration FO can be configured on different terminals.
Verify configuration.
Output level not compatible with
receiving device
See your transmitter installation manual. Verify
that the output level and the required receiving
input level are compatible.
Bad output circuit Perform loop test. See Section 5.3.
Incorrect internal/external power
configuration
Internal means that the transmitter will supply
power. External means that an external pull-up
resistor and source are required. Refer to your
transmitter installation manual for wiring. Verify
configuration is correct for desired application
(see Section 6.3).
Incorrect pulse width configuration Verify pulse width setting. See Section 6.6.3.
Output not powered (IS transmitters
only)
Check transmitter wiring. See Appendix B for
diagrams.
Table 10-2 Milliamp and frequency output problems and remedies (continued)
Symptom Possible cause Possible remedy