Transmitter Configuration and Use: Series 1000 and 2000 Transmitters 163
Calibrating the Transmitter continued
Enhanced Density Model 2500 CIOModel 1500 ANCalibration
Figure C-4 D3 or D3 and D4 density calibration – Communicator
Close shutoff valve
downstream from sensor
Fill sensor with D3 fluid
Diag/Service
Calibration
Density cal
Dens Pt 3
Enter density of D3 fluid
Perform Cal
Fill sensor with D4 fluid
HOME
Calibration in Progress
message
Density Calibration
Complete
message
On-Line Menu
OK
OK
Diag/Service
Calibration
Density cal
Dens Pt 4
Enter density of D4 fluid
Perform Cal
HOME
Calibration in Progress
message
Density Calibration
Complete
message
OK
OK
Exit