30 Transmitter Configuration and Use: Series 1000 and 2000 Transmitters
Flowmeter Startup continued
5.2 Applying power
Before you apply power to the flowmeter, close and tighten all housing covers.
Turn on the electrical power at the power supply. The flowmeter will automatically perform
diagnostic routines. When the flowmeter has completed its power-up sequence:
For Model 1700/2700 transmitters under normal conditions, the status LED on the display will
turn green and begin to flash,
For Model 1500/2500 transmitters under normal conditions, the status LED will turn green.
If the status LED exhibits different behavior, an alarm condition is present or transmitter zero
is in progress. See Section 7.6.
5.2.1 Communication methods after power-up
For Model 1700/2700 transmitters, all communication methods supported by the transmitter are
available immediately after power-up.
For Model 1500/2500 transmitters:
If you are using the Communicator, or ProLink II with HART protocol (HART/Bell202), you
can establish communication with the transmitter immediately after power-up, using terminals
21 and 22. See Chapter 4 for more information on using the Communicator, or Chapter 3 for
more information on using ProLink II.
If you are using ProLink II via the RS-485 physical layer, terminals 33 and 34 are available in
service port mode for 10 seconds immediately after power-up. If no connection is made during
this period, the terminals are automatically reset to the configured Modbus communication
parameters. Be sure to set the ProLink II connection parameters appropriately. See Chapter 3.
WARNING
Operating the flowmeter without covers in place
creates electrical hazards that can cause death,
injury, or property damage.
Make sure safety barrier partition and covers for the
field-wiring, circuit board compartments, electronics
module, and housing are all in place before applying
power to the transmitter.
WARNING
Using the service port to communicate with a Model
1700/2700 transmitter in a hazardous area can cause
an explosion.
Before using the service port to communicate with the
transmitter in a hazardous area, make sure the
atmosphere is free of explosive gases.