Transmitter Configuration and Use: Series 1000 and 2000 Transmitters 91
Optional Configuration continued
Required Configuration Optional ConfigurationUsing the TransmitterFlowmeter Startup
Example 3
Configuration:
Flow direction = Forward only
mA output: 4 mA = –100 g/s; 20 mA = 100 g/s
(See the first graph in Figure 8-2.)
As a result:
Under conditions of zero flow, the mA output is 12 mA.
Under conditions of forward flow, up to a flow rate of 100 g/s, the
mA output varies between 12 mA and 20 mA in proportion to (the
absolute value of) the flow rate.
Under conditions of forward flow, if (the absolute value of) the flow
rate equals or exceeds 100 g/s, the mA output is proportional to
the flow rate up to 20.5 mA, and will be level at 20.5 mA at higher
flow rates.
Under conditions of reverse flow, up to a flow rate of 100 g/s, the
mA output varies between 4 mA and 12 mA in inverse proportion
to the absolute value of the flow rate.
Under conditions of reverse flow, if the absolute value of the flow
rate equals or exceeds 100 g/s, the mA output is inversely
proportional to the flow rate down to 3.8 mA, and will be level at
3.8 mA at higher absolute values.
Table 8-6 Effect of flow direction on frequency outputs, totalizers, and digital communications
Flow direction value
Forward flow
(1)
(1) Process fluid flowing in same direction as flow direction arrow on sensor.
Frequency outputs Flow totals Flow values via digital comm.
Forward only Increase Increase Positive
Reverse only 0 Hz No change Positive
Bidirectional Increase Increase Positive
Absolute value Increase Increase Positive
(2)
(2) Refer to the digital communications status bits for an indication of whether flow is positive or negative.
Negate/Forward only Zero
(2)
No change Negative
Negate/Bidirectional Increase Decrease Negative
Flow direction value
Reverse flow
(3)
(3) Process fluid flowing in opposite direction from flow direction arrow on sensor.
Frequency outputs Flow totals Flow values via digital comm.
Forward only 0 Hz No change Negative
Reverse only Increase Increase Negative
Bidirectional Increase Decrease Negative
Absolute value Increase Increase Positive
(2)
Negate/Forward only Increase Increase Positive
Negate/Bidirectional Increase Increase Positive