37
Reference Manual
00809-0100-4803, Rev DA
Section 3: Configuration
April 2013
Configuration
Figure 3-19. Device - variable mapping tab
3.6.4 LCD display
The LCD features a four-line display and a 0-100% scaled bar graph. The first line of five
characters displays the output description, the second line of seven digits displays the actual
value, and the third line of six characters displays engineering units. The fourth line displays
“Error” when there is a problem detected with the transmitter. The LCD can also display
diagnostic messages. These diagnostic messages are listed in Table 5-1 on page 2.
The LCD tab allows the user to configure which variables will be shown on the LCD. Click the
check box next to each variable to select a variable for display. The transmitter will scroll through
the selected variables, showing each for three seconds.
Mass and Energy Flow Fast Keys 1, 3, 8
Direct Process Var. Output Fast Keys 1, 3, 8