4.4 Transmitter functional tests
Figure 4-6. Transmitter functional tests screen
4.4.1 Flow/energy calculation verification (Test calculation)
(Fully Compensated Mass and Energy Flow Feature Board only):
The Flow and Energy Calculation Verification Test allows the user to verify the flow configuration
of the 3051S MultiVariable Transmitter by entering expected values for the Differential Pressure,
Static Pressure and Process Temperature variables. Under the Flow/Energy Calculation Verification
heading, perform the following steps:
1. Click the Enable Test Calculation button.
2. Select Simulate DP option. Click Next.
3. Select DP Units from the drop-down menu. Click Next.
4. Enter the DP Value corresponding to the desired flow rate simulation. Click Next.
5. Repeat steps 1-3 for static pressure (Simulate AP/GP) and process temperature
(Simulate PT), if applicable.
6. Select View Results. Click Next. The simulated flow rate and corresponding flow
properties will be shown. Click Next.
7. Select Exit. Click Next. Leaving the Enable Test Calculation window automatically returns
all process variables fixed by the test calculation method to live process variable
measurements.
Mass and Energy Flow Fast Keys 1, 2, 3