Range and sensor limits
Range
3051SMV Differential Pressure Range and Sensor Limits
Minimum Span Range Limits
Ultra and Ultra for
Flow
Classic and Classic MV Upper (URL) Lower (LRL)
(1)
(1) Lower (LRL) is 0 inH
2
O (0 mbar) for Ultra for Flow.
0 0.1 inH
2
O (0,25 mbar) 0.1 inH
2
O (0,25 mbar) 3.0 inH
2
O (7,5 mbar) -3.0 inH
2
O (-7,5 mbar)
1 0.5 inH
2
O (1,24 mbar) 0.5 inH
2
O (1,24 mbar) 25.0 inH
2
O (62,3 mbar) -25.0 inH
2
O (-62,3 mbar)
2 1.3 inH
2
O (3,11 mbar) 2.5 inH
2
O (6,23 mbar) 250.0 inH
2
O (0,62 bar) -250.0 inH
2
O (-0,62 bar)
3 5.0 inH
2
O (12,4 mbar) 10.0 inH
2
O (24,9 mbar) 1000.0 inH
2
O (2,49 bar) -1000.0 inH
2
O (-2,49 bar)
4 1.5 psi (103,4 mbar) 3.0 psi (206,8 mbar) 300.0 psi (20,7 bar) -300.0 psi (-20,7 bar)
5 10.0 psi (689,5 mbar) 20.0 psi (1,38 bar) 2000.0 psi (137,9 bar) - 2000.0 psi (-137,9 bar)
Range
3051SMV Static Pressure Range and Sensor Limits
Minimum Span Range Limits
Ultra for Flow
Classic MV Upper (URL)
Lower (LRL)
(Absolute)
Lower (LRL)
(Gage)
(1)(2)
(1) Assumes atmospheric pressure of 14.7 psig (1 bar).
(2) Inert Fill: Minimum pressure = 1.5 psia (0,10 bar) or -13.2 psig (-0,91 bar).
3 4.0 psi (276 mbar) 8.0 psi (552 mbar) 800 psi (55,16 bar) 0.5 psia (34,5 mbar) -14.2 psig (-0,98 bar)
4 18.13 psi (1,25 bar) 36.26 psi (2,50 bar) 3626 psi (250.0 bar)
(3)
(3) For SP Range 4 and DP Range 1, the URL is 2000 psi (137,9 bar).
0.5 psia (34,5 mbar) -14.2 psig (-0,98 bar)
Process Temperature RTD Interface Range Limits
(1)
(1) Designed to accommodate a Pt 100 RTD sensor. Examples of compatible RTDs include Rosemount Series 68 and 78 RTD Temperature Sensors.
Minimum Span Upper (URL) Lower (LRL)
50 °F (28 °C) 1562 °F (850 °C) -328 °F (-200 °C)