Emerson 3051S Especificaciones Pagina 125

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 188
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 124
13
Reference Manual
00809-0100-4803, Rev DA
Section 5: Troubleshooting
April 2013
Troubleshooting
5.3 Measurement quality and limit status
The 3051S MultiVariable Transmitter is compliant with the HART Revision 6 Standard. One of
the most noticeable enhancements available with the HART 6 standard is that each variable has
a measurement quality and limit status. These statuses can be viewed in AMS, on a 475, or with
any HART 6 compatible host system. In AMS, variable statuses can be viewed by selecting
Variables in the upper left menu tree under the Configure/Setup heading.
Figure 5-1. Quality and limit status
Each variable status reading consists of two parts separated by a hyphen; Measurement Quality
and Limit Status.
Possible measurement quality readings
Good – Displayed during normal device operation.
Poor Accuracy – Indicates the accuracy of the variable measurement has been compromised.
Example: The module temperature sensor failed and is no longer compensating the differential
pressure and status pressure measurements.
Bad – Indicates the variable has failed. Example: A differential pressure, static pressure, or
process temperature sensor failure.
Possible limit status readings
Not Limited – Displayed during normal device operation.
High Limited – Indicates the current variable reading has gone above the transmitter’s maximum
possible reading and is no longer representative of the actual variable measurement.
Measurement Quality and Limit Status
Vista de pagina 124
1 2 ... 120 121 122 123 124 125 126 127 128 129 130 ... 187 188

Comentarios a estos manuales

Sin comentarios