3-2
EN
Rosemount Inc
Surges and Transients
To protect against high-energy transients, install the transmitter into a suitable
connection head with the Rosemount Model 470 Transient Protector. Refer to the
Model 470 Transient Protector Product Data Sheet (document number 00813-0100-
4191) for more information.
CONFIGURATION
Online Menu
The HART Communicator Online Menu provides direct access to Fast Key
Sequences and all software functions of the transmitter. it appears automatically if
the communicator is connected to an active loop with an operating transmitter or it
can be selected from the Main Menu.
Hart Fast Key Feature
HART Fast Keys are operational only from the Online Menu. To use them, you
must return to the Online Menu by pressing HOME (F3) when it is available. If you
do not start at the Online Menu, the HART Fast Key Sequences will not function
properly.
HART Fast Key Description Example: Changing the Date
Hart Fast Key sequences are made up of a
series of number that correspond to the
individual options in each step of the menu
structure. They can be identified using the
following conventions:
1 through 9 - Refer to the keys on the
alpha-numeric keypad located below the
dedicated keypad.
1. Following the menu structure, press 1 to
reach Device Setup.
2. Press 3 for Configuration.
3. Press 4 for Device information.
4. Press 2 for Date.
The Corresponding HART Fast Key sequence
is 1, 3, 4, 2.