
Reference Manual
00809-0100-4801, Rev FA
October 2010
4-11
Rosemount 3051S Series
The following specifications show the static pressure effect for the 3051S
Range 4 and Range 5 transmitters used in differential pressure applications:
Zero Effect:
± 0.1% of the upper range limit per 1000 psi (69 bar) for line pressures
from 0 to 2000 psi (0 to 138 bar)
For line pressures above 2000 psi (138 bar), the zero effect error is ± 0.2%
of the upper range limit plus an additional ± 0.2% of upper range limit error
for each 1000 psi (69 bar) of line pressure above 2000 psi (138 bar).
Example: Line pressure is 3000 psi (207 bar). Zero effect error calculation:
± {0.2 + 0.2 x [3 kpsi - 2 kpsi]} = ± 0.4% of the upper range limit
Span Effect:
Correctable to ±0.2% of reading per 1000 psi (69 bar) for line pressures
from 0 to 3626 psi (0 to 250 bar)
The systematic span shift caused by the application of static line pressure is
-1.00% of reading per 1000 psi (69 bar) for Range 4 transmitters, and -1.25%
of reading per 1000 psi (69 bar) for Range 5 transmitters.
Use the following example to compute corrected input values.
Example
A transmitter with model number 3051S_CD4 will be used in a differential
pressure application where the static line pressure is 1200 psi (83 bar).
The transmitter output is ranged with 4 mA at 500 inH
2
O (1,2 bar) and
20 mA at 1500 inH
2
O (3,7 bar).
To correct for systematic error caused by high static line pressure, first use
the following formulas to determine corrected values for the low trim and
high trim.
LT = LRV + S x (LRV) x P
HT = URV + S x (URV) x P
Where: LT = Corrected Low Trim Value
LRV = Lower Range Value
S = –(Span shift per specification)
P = Static Line Pressure
Where: HT = Corrected High Trim Value
URV = Upper Range Value
S = –(Span shift per specification)
P = Static Line Pressure
Comentarios a estos manuales