NOTE
If the transmitter security jumper/switch is ON, adjustments to the zero and
span will not be able to be made. Refer to “Configure Security and Alarm” on
page 2-13 for security information.
Rerange with a Field Communicator v3.3 Only
The easiest and most popular way to rerange is to use the Field
Communicator only. This method changes the values of the analog 4 and 20
mA points independently without a pressure input.
From the HOME screen, enter the fast key sequence “Rerange with a
Communicator Only.”
1. At “Keypad Input” select 1 and use the keypad to enter lower range
value.
2. From “Keypad Input” select 2 and use the keypad to enter upper range
value.
Rerange with a Pressure Input Source and Field Communicator v3.3
Reranging using the Field Communicator and a pressure source or process
pressure is a way of reranging the transmitter when specific 4 and 20 mA
points are unknown.
NOTE
The span is maintained when the 4 mA point is set. The span changes when
the 20 mA point is set. If the lower range point is set to a value that causes the
upper range point to exceed the sensor limit, the upper range point is
automatically set to the sensor limit, and the span is adjusted accordingly.
1. From the HOME screen, enter the fast key sequence “Rerange with a
Pressure Input Source and a Field Communicator” to configure lower
and upper range values and follow the on-line instructions.
Traditional Fast Keys
1, 2, 3, 1, 1
Device Dashboard Fast Keys
1, 5
Traditional Fast Keys
1, 2, 3, 1, 2
Device Dashboard Fast Keys
3, 4, 1