Reference Manual
00809-0100-4801, Rev FA
October 2010
7-21
Rosemount 3051S Series
Overview The transmitter is shipped with Power Advisory off as default and without any
loop characterization performed. Once the transmitter is installed and
powered up, loop characterization must be performed for Power Advisory
diagnostic to function.
When the user initiates a loop characterization, the transmitter will check to
see if the loop has sufficient power for proper operation. Then the transmitter
will drive the analog output to both 4mA and 20mA to establish a baseline and
determine the maximum allowable terminal voltage deviation. Once this is
complete, the user enters a sensitivity threshold called “Terminal Voltage
Deviation Limit” and a check is in place to make sure this threshold value is
valid.
Once the loop has been characterized and Terminal Voltage Deviation Limit is
set, Power Advisory actively monitors the electrical loop for deviations from
the baseline. If the terminal voltage has changed relative to the expected
baseline value, exceeding the configured Terminal Voltage Deviation Limit,
the transmitter can generate an alert or alarm.
NOTE
Power Advisory diagnostic in the Rosemount 3051S HART pressure
transmitter monitors and detects changes in the terminal voltage from
expected values to detect common failures. It is not possible to predict and
detect all types of electrical failures on the 4-20mA output. Therefore,
Rosemount cannot absolutely warrant or guarantee that Power Advisory
Diagnostic will accurately detect failures under all circumstances.
Configuration For inexperienced users, guided setup is recommended. Guided setup walks
the user through settings that configure the Power Advisory diagnostic for
most common usage and applications.
Figure 7-17. Guided Setup
Menu
The rest of the configuration section explains the parameters for manual
configuration of Power Advisory diagnostic.
Device Dashboard
Fast Keys
2, 1, 2, 2