Emerson ROSEMOUNT 5300 Especificaciones Pagina 52

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 106
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 51
Reference Manual
00809-0600-4026, Rev BA
December 2011
Rosemount Process Radar
4-8
Figure 4-8. Echo Curve plot
indicating that the amplitude
threshold for the interface peak
is too high
If the Interface Threshold is too high, the signal amplitude peak at the
interface between the upper and lower products is not detected.
Recommended Actions
Lower Interface Threshold.
Strong fixed echo - Bent probe / Probe contacting Nozzle
A bent probe in contact with metal creates a strong echo and
measurement will lock onto the strong echo. Note that contamination also
may create strong disturbance echoes.
Figure 4-9. Measurement data
using a bent probe from refining
best practices
3.0 5.0
The Interface Threshold is above
the interface peak
Amplitude, mV
Distance, m
Interface Threshold
Surface Threshold
Reference Threshold
Vista de pagina 51
1 2 ... 47 48 49 50 51 52 53 54 55 56 57 ... 105 106

Comentarios a estos manuales

Sin comentarios