Reference Manual
00809-0100-4102, Rev AA
Section 3: Installation
May 2013
Installation
Figure 3-12. Inline gage low side pressure port
A. Low side pressure port (atmospheric reference)
a
3.4.5 Power Module installation
Figure 3-13. Power Module
A. Power Module (5/64-inch hex wrench required)
Do not apply torque directly to the sensor module. Rotation between the sensor module
and the process connection can damage the electronics. To avoid damage, apply torque
only to the hex-shaped process connection.
A
Sensor Module
Process Connection
A