6 Micro Motion
®
ELITE
®
Flow and Density Meters
Gas flow performance
When selecting sensors for gas applications, measurement accuracy is a function of fluid mass flow rate independent of operating
temperature, pressure, or composition. However, pressure drop through the sensor is dependent upon operating temperature,
pressure, and fluid composition. Therefore, when selecting a sensor for any particular gas application, it is highly recommended that
each sensor be sized using the Micro Motion product selector, available at www.micromotion.com
.
Mass Volume
(1)
(1) Standard (SCFM) reference conditions are 14.7 psia and 68 °F. Normal (Nm
3
/h) reference conditions are 1.013 bar and 0 °C.
lb/min kg/h SCFM Nm
3
/h
Flow rates that produce
approximately 10 psi (0.68 bar)
pressure drop on air
(2)
(2) Air at 68 °F (20 °C) and 100 psia (6.8 bar).
CMFS010 0.3 8 4 6
CMFS015 1 24 12 18
CMF010M, H 0.30 8 4 6
CMF010P 0.2 6 3 5
CMF025 4 110 60 90
CMF050 10 300 145 230
CMF100 50 1300 640 1000
CMF200 150 4000 2000 3100
CMF300 490 13,300 6500 10,300
CMF400 1250 34,000 16,600 26,250
Flow rates that produce
approximately 50 psi (3.4 bar)
pressure drop on natural gas
(3)
(3) Natural gas with MW 16.675 at 68 °F (20 °C) and 500 psia (34.0 bar).
CMFS010 1 30 30 45
CMFS015 3 90 90 130
CMF010M, H 1 30 30 45
CMF010P 0.9 25 20 35
CMF025 16 450 380 600
CMF050 40 1140 970 1530
CMF100 185 5000 4300 6700
CMF200 560 15,200 13,000 20,500
CMF300 1850 50,500 43,000 68,000
CMF400 4700 128,000 109,000 172,000