Micro Motion
®
ELITE
®
Flow and Density Meters 5
Liquid flow performance
Mass Volume
(1)
(1) Specifications for volumetric flow rate are based on a process-fluid density of 1 g/cm
3
(1000 kg/m
3
). For fluids with density other than
1g/cm
3
(1000 kg/m
3
), the volumetric flow rate equals the mass flow rate divided by the fluid’s density.
lb/min kg/h gal/min l/h bbl/h m
3
/h
Maximum flow rate CMFS010 4 108 0.5 108
CMFS015 12 330 1.5 330
CMF010 4 108 0.5 108
CMF025 80 2180 10 2180
CMF050 250 6800 30 6800
CMF100 1000 27,200 120 27,200
CMF200 3200 87,100 385 87,100 550 87
CMF300 10,000 272,000 1200 272,000 1700 272
CMF400 20,000 545,000 2400 545,000 3400 545
Typical accuracy, turndown, and pressure drop with CMF100 and 2400S or enhanced core processor
The graph below is an example of the relationship between accuracy, turndown, and pressure drop when measuring the flow of
water with a Model CMF100 sensor and Model 2400S transmitter or enhanced core processor.
Actual pressure drop is dependent on process conditions. To determine accuracy, turndown, and pressure drop with your process
variables, use the Micro Motion product selector, available at www.micromotion.com
.
Turndown from maximum flow rate 500:1 100:1 20:1 10:1 2:1
Accuracy ±% 1.25 0.25 0.05 0.05 0.05
Pressure drop psi ~0 ~0 0.2 0.7 13.5
bar ~0 ~0 0.01 0.05 0.93