MODEL 6081-P pH/ORP SECTION 8.0
PROGRAMMING THE TRANSMITTER
8.6 MAKING HART RELATED SETTINGS
For more information refer to Sec. 1.0
8.7 RESETTING FACTORY CALIBRATION AND FACTORY DEFAULT SETTINGS
8.7.1 Purpose
This section describes how to install factory calibration and default values. The process also clears all fault
messages and returns the display to the first quick start screen.
8.7.2 Procedure: Installing default settings.
1. Press MENU. The menu screen appears. Choose Program.
2. Choose >>.
3. Choose >>.
4. Choose ResetTransmitter.
5. Choose Yes or No. Choosing Yes clears previous settings and calibrations and returns the transmitter to the
first quick start screen.