Reference Manual
MHM-97408, Rev. 2
June 2008
CSI 9420
3-8
Figure 3-7. Mounting
illustrations for right angle Quick
Connect sensors.
Secure Cabling 1. Secure the sensor cable to the machine approximately 4 to 5 inches
from the mounting location using an appropriate size cable clamp. Do
not curl into a bending radius of less than 2.8 inches. Refer to
Figure 3-8 for an illustration.
2. If pulling cable is not currently scheduled, it will be necessary to
secure the bundled sensor cables in such a manner that no strain is
placed on the integral sensor/cable connectors. Do not allow bundled
cable to hang from the sensors. Prevent damage to exposed cable
such as on plant floors, maintenance access areas, footholds and the
like.
Figure 3-8. Illustration of a
secure cable with a temporary
cable anchor.