8 Micro Motion
®
Series 3000 Transmitters and Controllers
Digital communications
RS-485 One pair of terminals supports SP (service port) mode or RS-485 mode. On device power-up, the
user has 10 seconds to connect in SP mode. After 10 seconds, the terminals switch to RS-485
mode. The communication parameters in SP mode are:
Protocol Modbus RTU
Data rate 38,400 baud
Parity No parity
Stop bit One stop bit
Address 111
In RS-485 mode, the Series 3000 can communicate via Modbus RTU, Modbus ASCII, or HART
(1)
protocols. The communication parameters are configurable with ProLink III software, Modbus, or
the display. The shipped default parameters are:
(1) HART protocol supports configuration of basic transmitter functionality, configuration of the petroleum measurement application, and
reading/writing batch setpoints and totals. HART protocol does not support configuration of the batch, concentration measurement, or
NOC applications. Both Modbus and the display provide configuration and control of all transmitter functionality.
Data rate 9600 baud
Parity Odd
Stop bit One stop bit
The RS-485 port can also be configured as a printer port (see below).
HART Bell 202
(1)
The HART Bell 202 signal is superimposed on the primary mA output, and is available for host
system interface or for polling external measurement devices. The HART Bell 202 communication
parameters are:
Frequency 1.2 and 2.2 KHz
Amplitude 0.8 mA peak-to-peak
Data rate 1200 baud
Resistance Loop requires 250 to 600 ohms resistance
Printer port The RS-485 port can be configured as a printer port. When the RS-485 port is
configured as a printer port, it cannot be used for any other function.
Requires external RS-232 adapter (not included)
With Epson TM-U295 ticket printer, detects and reports “paper out” conditions.