MODEL Xmt-P SECTION 10.0
TROUBLESHOOTING
10.9 MEASURING REFERENCE VOLTAGE
Some processes contain substances that poison or shift the poten-
tial of the reference electrode. Sulfide is a good example.
Prolonged exposure to sulfide converts the reference electrode
from a silver/silver chloride electrode to a silver/silver sulfide elec-
trode. The change in reference voltage is several hundred millivolts.
A good way to check for poisoning is to compare the voltage of the
reference electrode with a silver/silver chloride electrode known to
be good. The reference electrode from a new sensor is best. See
Figure 10-4. If the reference electrode is good, the voltage differ-
ence should be no more than about 20 mV. A poisoned reference
electrode usually requires replacement.
FIGURE 10-4. Checking for a Poisoned
Reference Electrode.
Refer to the sensor wiring diagram to identify the
reference leads. A laboratory silver/silver chloride
electrode can be used in place of the second sensor.